Cabling Requirements for G0 Target Controls in Hall C 25-may-2001 Conductors required from NEAR TARGET to near racks in hall: there are a total of 6 10-pin feedthroughs on the target that have signals going over to the electronics racks. About 45 conductors on the feedthroughs are used. There will be a few additional signals related to interlocks and vacuum readout, so assuming about 60 conductors is reasonable. 1) Temperature sensors (4-wire resistance measurement) # conductors: 24 shielded in groups of 4 signal type: mV DC signals 2) Tachometer # conductors: 2 conductors + shield signal type: TTL pulse train 3) Low power heater # conductors: 2 signal type: DC, up to 40 V, 2 Amps 4) High power heater # conductors: 6 signal type: DC, up to 40 V, 10 Amps 5) Pump motor # conductors: 3 signal type: (?: ask Silviu) *** note this one could introduce electrical noise. 6) Vacuum: # conductors: about 10 signal type: DC (24 V max), low current 7) Position: a) control: # conductors: ?? signal type: 120 V, 0.5 Amps b) readout: # conductors: 8 (using 4 BNC cables at present) signal type: DC up to 10 V, low current c) microswitches: # conductors: 8 signal type: DC, closes a contact 9) JT valve: # conductors: ?? signal type: low voltage and current