Experiment I: Measurement of the Speed of
Light 
  - hints on laser alignment
- Don't look through the eyepiece unless the rotating mirror is ON.
- There is significant backlash in the vernier - think about how to take all your data
    while turning the vernier only in one direction.
Experiment II: Experimental Atomic
Spectroscopy 
  - Optimizing the slit widths for the best resolution is best done with the H/D source.
- Turn the flourescent lights off when you are making a scan to cut down on the electrical
    noise coming from the lights.
- Don't forget to calibrate the display on the microprocessor against the dial indicator
    on the monochrometer. If you don't, you may run the monochrometer into either its upper or
    lower limit. If this happens, contact one of the instructors right away so it can be
    reset.
Experiment III: The Franck-Hertz Experiment
 
  - Give yourself a quick refresher on RC circuits for oscillating signals before connecting
    the electronics.
- REMOVE THE RC FILTER WHEN YOU CALIBRATE THE 40V POWER SUPPLY!
- Allen suggests turning off the temperature controller just before taking a scan with
    ACQUIRE (don't forget to turn it back on though.).
- Simultaneously recording the Franck-Hertz tube current in channel 0 and the ramping
    voltage in channel 1 is tricky. The ramping power supply is floated away from ground by
    the retarding voltage.  I suggest measuring from ground to the filament, this
    measures the sum of the retarding voltage and the ramp voltage.
- Think hard about how to fit the peaks, and how to "normalize" the curve to
    correct for the fact that the overall current rises with the accelerating voltage.
Experiment IV: Determining Planck's Constant
  - Take a look at the plot of intensity vs
    wavelength to find the best peaks to look at.
- Give yourself a quick refresher on RC circuits for oscillating signals before connecting
    the electronics.
- You might be interested in taking a look at this recent measurement of
    Planck's constant at NIST in Gaithersburg, MD. See the associated article (21 Sep 98)
    by E.R. Williams, et al., Phys.
    Rev. Lett.  Vol 81, (1998) 2404.
- Look at Melissinos (p. 24) for suggestions on how to analyze the data to find the
    stopping voltage.  Think about this - other methods may work even better!
Experiment V: 
Experiment VI: Measurement of (e/m) for the
electron 
  - You MUST go over the operation of the vacuum system with either Allen Monroe, Tom
    Baldwin, or one of the instructors before starting the experiment! See also Appendix A in
    the lab manual.
-  Here's a 
    calibration from 1996, plus a plot showing the residual
    errors.  Refitting this data (2003) shows the uncertainties:  intercept tof
    (0.250 +/-0.009) G and the slope of (47.3 +/- 0.1) G/Amp.  This whole calibration
    needs to be redone with a stable series resistor for measuring the current.
- Think about the contribution of the earth's field to your measurement, and how you can
    account for (or get rid of) it.
- See the updated instructions for operating the vacuum
    system, replaces Section III of Appendix A.
Experiment VII: Introduction to Nuclear
Spectroscopy I 
  - NOTE: this experiment along with VIII and IX use a data acquisition and software
    package  specifically designed for nuclear spectroscopy. You should take a little
    time to familiarize yourself with the program before attempting to take data. See  Appendix C
    for a brief description of the program.
- For this experiment, and also exps VIII and IX, here is a brief description on using
    either the Tektronix
    2245A or the digital Tektronix
    TDS360 scopes.
Experiment VIII: Introduction to Nuclear
Spectroscopy II 
  - See the notes in exp VII regarding data acquisition.
- You can find more and up-to-date information about the particular nuclei you are
    studying at the National Nuclear Data Center at
    Brookhaven National Lab.  This picture  of the 56Mn->56Fe
    gamma ray scheme was generated using the MIRD database. Also, take a look at the 228Th
    spectrum hanging on the wall near the Exp 8 setup for your energy calibration.
- For this experiment, and also exps VII and IX, here is a brief description on using
    either the Tektronix
    2245A or the digital Tektronix
    TDS360 scopes.
Experiment IX: Gamma-Gamma Angular Correlation
  - See the notes in exp VII regarding data acquisition.
- Use the fast oscilloscope (Tektronix
    2245A or Tektronix
    TDS360) from exps VII/VIII or exp XI to check the coincidence timing! Follow the
    signals all the way through the circuit to make sure the signals going into the
    coincidence circuit are properly timed.
- There is a stronger 60Co source inside the little lead can in the lead-lined
    box. The metal rod (not the block) is the active area of the source. Calculate for
    yourself how much stronger this source is than the one in the wooden box.
- The detection efficiency of the two-detector system changes as a function of angle. This
    affects the overall shape of the data. The effect can be estimated using Monte Carlo
    techniques. Run the Monte Carlo Program called MC.EXE(written
    by Prof. CC Chang)  to see the effect. It is located in the DOS directory of programs
    on the lab computers.
- You will need to fit your data to either Legendre Polynomials, or to equation IX-1 in
    the lab manual. A sample fitting program, using EXCEL, can be found at p:\p405\Analysis Tools\Excel Tools\ParaFit-mc.xls. This is a
    parabolic fit with 3 parameters. The data have been entered such that the independent
    variable is cos^2(theta). It's rather clumsy, but functional.
Experiment X: The Hall Effect in Metals
  - You MUST go over the vacuum system with either Allen Monroe, Tom Baldwin, or one of the
    instructors before starting the experiment! See also Appendix A.
- You will need to take a look at Appendix E
    on sample preparation, which is not in the lab manual.
- See one of the instructors regarding thickness measurements: there is a relatively new
    apparatus donated by Dr. Webb.  A manual for it can be found in this Appendix D.
Experiment XI: Cosmic Rays 
  - To get the solid angle of the detector system, you have to do a four-dimensional
    integral: you can either use Mathematica, or you can look for the nifty little pascal
    program written by Prof. C.C. Chang, called INTEGRAL.EXE,
    which is located on the P405 PCs in  p:\p405\exp11\ .
- Figure XI-2 in the lab manual is OK, but not quite accurate. This one
    is a little better.
- Use the digital oscilloscope (Tektronix
    TDS360) to check the coincidence timing.
- BE SURE THE HV SUPPLY is set to NEGATIVE (-)!
- Further information on the various components of cosmic rays can be found in the
    Particle Data Group's Review of
    Particle Properties.  Look under Astrophysics and Cosmology, then under Cosmic
    Rays.